HAMILTON-C1/T1/MR1 | How to perform preoperational checks SW 2.x

Аватар автора
Вентиляция: заботьтесь о качестве воздуха в доме
In this video, we show you how to perform the preop checks HAMILTON-C1/T1/MR1 ventilators for SW 2.x. It is demonstrated on the HAMILTON-T1 but applies accordingly on the HAMILTON-C1 and the HAMILTON-MR1. We are passionate about investing in the growth and development of our customers and the wider ventilation community. That is why we offer free and open e-learning modules on mechanical ventilation and ventilators through our e-Academy. 00:00 Intro 00:48 Before starting 01:06 Turning on the ventilator 01:24 Accessing tests and calibration functions 02:03 Performing the tightness test 02:45 Calibrating the flow sensor 03:40 Zero calibrating the CO2 sensor and airway adapter 04:47 Preoperational checks with a coaxial breathing circuit 05:07 Calibrating the O2 sensor with High Pressure Oxygen (HPO) 06:09 Calibrating the O2 sensor (with LPO or HPO with less than 99% O2) 06:31 Starting ventilation or turning off the ventilator Next step: How to start ventilation